Innovative solutions for electrochemical research
OctoStat200

OctoStat200.dc

Choose configuration:

Each channel: ±200mA/±10V

8-channel Potentiostat/Galvanostat/ZRA

Grounded CE configuration allowing multiple channels in one cell

19" rack-mountable housing

You can buy this product from your local reseller:

Ivium Technologies B.V.

Request information

Product description

Dedicated application

The 8-channel instrument is designed for applications where multiple cells are operated in parallel for long periods of time, and is capable of all standard DC electrochemical techniques.

The OctoStat200.dc has a grounded CE configuration which allows:

  • multiple channels in 1 cell (bipotentiostat, WE-arrays, Devanathan, etc.)
  • sharing CE and RE for multiple cells or samples
  • combination with an other potentiostat to form a bipotentiostat

Key specifications 
– Scan range: ±10V/±200mA
– Grounded CE configuration
– Current ranges from 10nA to 100mA
– Max. acquisition rate: 1,000 pnts/s

Complete solution

The OctoStat offers a complete package. The hardware includes low current measurement possibility down to 10nA full-scale with sub-fA resolution, and analog measurement channels for secondary signals such as temperature. All measurement and dataprocessing software is included.

Safety

The maximum current or potential of the instrument can be defined by the operator. Thus valuable samples may be protected, and unsafe situations prevented.

Data security & Connection

The OctoStat is equipped with an integrated DataSecure module:

  • Data is stored on board independent of the computer.
  • Connection of multiple PCs is possible.

Hardware

  • Automatic current range selection: 10nA to 100mA
  • Simultaneous acquisition of current/potential and peripheral analog signal
  • Real time data up to 1,000 pnts/s
  • Applicable scanrates: 1µV/s to 10,000V/s
  • Minimum interval time 1ms

Software

The OctoStat200.dc is controlled via Ivium’s own software, IviumSoft. This versatile, yet intuitive software, allows instrument control, data management and analysis, etc. The OctoStat can also be controlled from LabVIEWTM, C++, Delphi, etc. A full suite of software is included as standard with each instrument.

  • Control of separate channels individually
  • All channels simultaneously
  • Synchronous start
  • Data from different channels plotted individually or in 1 single graph
Read the full description

Specifications

System performance
Current compliance
±200mA
Maximum compliance voltage
±10V
Maximum applied voltage
±10V
Electrode connections
4; WE, CE, RE, S (and GND) with 4mm banana plugs
Potentiostat bandwidth
>1kHz
Signal acquisition
Dual channel 24bit ADC, 1,000 samples/s
Potentiostat
Applied potential range
±10V, with 0.33mV res.
Applied potential accuracy
0.2%, or 2mV
Current ranges
±10nA to ±100mA
Measured current resolution
0.0005% of current range, minimum 0.5fA
Measured current accuracy
20pA + 0.025% of FSR
Galvanostat
Applied current resolution
0.012% of applied current range
Applied current accuracy
0.2%
Galvanostatic current ranges
±10nA to ±100mA
Potential ranges
±10V
Measured potential resolution
1µV
Measured potential accuracy
0.2% or 2mV
Electrometer
Input impedance
>1000Gohm //<10pF
Input bias current
<20pA
Bandwidth
>100kHz
Peripheral connections
Analog in/out
2 per channel (±10V input only, 1000GOhm, 24bit resolution)
Data Acquisition
Data acquisition time
1ms min.
Stored no. of datapoints
20M each channel
Environment
Interfacing
USB/LAN (Ethernet)
PC requirements
Windows 8/10/11, with free USB port
Module height
1U
Integrated DataSecure
Data acquisition time: 2ms minimum, Stored no. of data points: 20M each channel
Specifications subject to change, Ivium Technologies ©2025

Model comparison

OctoStat30
OctoStat200
OctoStat200.dc
OctoStat5000
OctoBoost16000
System performance
Current compliance
±30mA
±200mA
±200mA
±5A
±16A
Maximum compliance voltage
±10V
±10V
±10V
±10V
-2 to +9V, or ±5V (user selectable)
Maximum applied voltage
±10V
±10V
±10V
±10V
-2 to +9V, or ±5V (user selectable)
Electrode connections
4; WE, CE, RE, S (and GND) with 4mm banana plugs
4; WE, CE, RE, S (and GND) with 4mm banana plugs
4; WE, CE, RE, S (and GND) with 4mm banana plugs
4; WE, CE, RE, S (and GND) with 4mm banana plugs
-
Potentiostat bandwidth
>500kHz
>500kHz
>1kHz
>500kHz
-
Stability settings
High Speed, Standard and High Stability
High Speed, Standard and High Stability
-
High Speed, Standard and High Stability
-
Programmable response filter
1MHz, 100kHz, 10kHz, 1kHz, 10Hz
1MHz, 100kHz, 10kHz, 1kHz, 10Hz
-
1MHz, 100kHz, 10kHz, 1kHz, 10Hz
-
Signal acquisition
Dual channel 18bit ADC, 300,000 samples/s stored in instrument memory (max. 1M datapoints)
Dual channel 18bit ADC, 300,000 samples/s stored in instrument memory (max. 1M datapoints)
Dual channel 24bit ADC, 1,000 samples/s
Dual channel 18bit ADC, 300,000 samples/s stored in instrument memory (max. 1M datapoints)
-
Channel combination
No
No
-
No
Yes
Channel combination configuration
-
-
-
-
Neighbouring channels combinable, e.g. 4 x ±32A, 2 x ±64A, 1 x ±64A and 4 x ±16A, 1 x ±128A, etc.
Potentiostat
Applied potential range
±10V, 0.08mV res.
±10V, 0.08mV res.
±10V, with 0.33mV res.
±10V, 0.08mV res.
-
Applied potential accuracy
0.2%, or 2mV
0.2%, or 2mV
0.2%, or 2mV
0.2%, or 2mV
-
Current ranges
±100pA to ±10mA in 9 decades
±100pA to ±100mA in 10 decades
±10nA to ±100mA
±100pA to ±10A
±10A and ±100A
Measured current resolution
0.003% of current range, minimum 3fA
0.003% of current range, minimum 3fA
0.0005% of current range, minimum 0.5fA
0.003% of current range, minimum 3fA
-
Measured current accuracy
20pA + 0.025% of FSR
20pA + 0.025% of FSR
20pA + 0.025% of FSR
20pA + 0.025% of FSR
0.025% of FSR
Galvanostat
Applied current resolution
0.008% of applied current range
0.008% of applied current range
0.012% of applied current range
0.008% of applied current range
-
Applied current accuracy
0.2%
0.2%
0.2%
0.2%
-
Galvanostatic current ranges
±10nA to ±10mA
±10nA to ±100mA
±10nA to ±100mA
±10nA to ±10A
-
Potential ranges
±1mV, ±4mV, ±10mV, ±40mV, ±100mV, ±400mV, ±1V, ±4V, ±10V
±1mV, ±4mV, ±10mV, ±40mV, ±100mV, ±400mV, ±1V, ±4V, ±10V
±10V
±1mV, ±4mV, ±10mV, ±40mV, ±100mV, ±400mV, ±1V, ±4V, ±10V
-
Measured potential resolution
0.0008% of potential range, minimum 7nV
0.0008% of potential range, minimum 7nV
1µV
0.0008% of potential range, minimum 7nV
-
Measured potential accuracy
0.2% or 2mV
0.2% or 2mV
0.2% or 2mV
0.2% or 2mV
-
Impedance Analyser
Frequency range
10µHz to 100kHz (1MHz optional)
10µHz to 100kHz (1MHz optional)
-
10µHz to 100kHz (1MHz optional)
10µHz to 10kHz
Amplitude
0.15mV to 2.0V, or 0.03% to 100% of current range
0.15mV to 2.0V, or 0.03% to 100% of current range
-
0.15mV to 2.0V, or 0.03% to 100% of current range
0.15mV to 2.0V potentiostatic, or current compliance of the (combination of) channels galvanostatic
DC offset
16bit DC offset subtraction and 2 DC-decoupling filters
16bit DC offset subtraction and 2 DC-decoupling filters
-
16bit DC offset subtraction and 2 DC-decoupling filters
-
Electrometer
Input impedance
>1000Gohm //<10pF
>1000Gohm //<10pF
>1000Gohm //<10pF
>1000Gohm //<10pF
-
Input bias current
<20pA
<20pA
<20pA
<20pA
-
Bandwidth
>5MHz
>5MHz
>100kHz
>5MHz
-
Special functions
Ohmic drop compensation
2V/current range
2V/current range
-
2V/current range
-
Peripheral connections
Analog in/out
2 per channel (±10V, inputs 18bit resolution, output 16bit resolution, bandwidth 40kHz)
2 per channel (±10V, inputs 18bit resolution, output 16bit resolution, bandwidth 40kHz)
2 per channel (±10V input only, 1000GOhm, 24bit resolution)
2 per channel (±10V, inputs 18bit resolution, output 16bit resolution, bandwidth 40kHz)
-
Data Acquisition
Data acquisition time
3.33µs min.
3.33µs min.
1ms min.
3.33µs min.
-
Stored no. of datapoints
20M each channel
20M each channel
20M each channel
20M each channel
-
Environment
Interfacing
USB/LAN (Ethernet)
USB/LAN (Ethernet)
USB/LAN (Ethernet)
USB/LAN (Ethernet)
-
PC requirements
Windows 8/10/11, with free USB port
Windows 8/10/11, with free USB port
Windows 8/10/11, with free USB port
Windows 8/10/11, with free USB port
-
Module height
1U
1U
1U
2U
3U
Integrated DataSecure
Data acquisition time: 2ms minimum, Stored no. of data points: 20M each channel
Data acquisition time: 2ms minimum, Stored no. of data points: 20M each channel
Data acquisition time: 2ms minimum, Stored no. of data points: 20M each channel
Data acquisition time: 2ms minimum, Stored no. of data points: 20M each channel
-

Need help?

Our experts are ready to answer your questions.

Information request

Postal address

  • Ivium Technologies B.V.
  • De Zaale 11
  • 5612 AJ Eindhoven
  • The Netherlands

Visiting address

  • Ivium Technologies B.V.
  • De Lismortel 31
  • 5612 AR Eindhoven
  • The Netherlands

Contact

  • T +31 40 2390600
  • F
  • E info@ivium.com
©2025 Ivium Technologies B.V.

Sitemap