Innovative solutions for electrochemical research
HE80

HE80-1Q

Choose configuration:

Current compliance: 0 to +80A

Voltage compliance: +0.5 to +20V

FRA/EIS: 10µHz-10kHz

Buy from your local reseller (US):

Scribner, LLC

Request information

Product description

Potentiostat
The HE80-1Q is a single-quadrant potentiostat with integrated FRA/EIS. It is a unipolar device capable of +0.5 to +20V, and 0 to +80A, and has been specifically designed with electrolysers/hydrogen generation, battery charging, and electrodeposition in mind.

Key specifications
– Scan range: +0.5 to +20V/0 to +80A
– 4 Electrode connection: WE/CE by M8 screw terminals, RE/S by 4mm bananas
– Max. acquisition rate: 300,000 points/s
– Size: 44 x 37 x 8.8cm
– Weight: 15kg

Hardware

  • Real time data up to 500 pnts/s. Acquisition up to 300,000 pnts/s is stored in instrument memory (1M sample data buffer)
  • Minimum interval time: 3.33µs (0.125µs resolution)

Software:
The HE80-1Q is controlled via Ivium’s own IviumSoft. This versatile, yet intuitive software package allows instrument control, data management and analysis, etc. IviumSoft can also interface with and be controlled from LabVIEWTM, C++, Delphi, etc. A full suite of IviumSoft is included with each instrument as standard.

Read the full description

Specifications

System performance
Current compliance
0 to +80A
Maximum compliance voltage
+20V
Maximum applied voltage
+0.5 to +20V
Electrode connections
4; WE and CE with M8 screw connectors, RE and S with 4mm bananas
Potentiostat bandwidth
>5kHz
Stability settings
Standard and High Stability
Programmable response filter
1MHz, 100kHz, 10kHz, 1kHz, 10Hz
Signal acquisition
Dual channel 18bit ADC, 300,000 samples/s stored in instrument memory (max. 1M datapoints)
Potentiostat
Applied potential range
+0.5 to +20V, 0.15mV res.
Applied potential accuracy
0.2%, or 2mV
Current ranges
+10A and +100A
Measured current resolution
0.003% of current range, minimum 0.3mA
Measured current accuracy
0.1% of range + 0.2% of value
Galvanostat
Applied current resolution
0.008% of applied current range
Applied current accuracy
0.2%
Galvanostatic current ranges
+10A to +100A
Potential ranges
±2mV, ±8mV, ±20mV, ±80mV, ±200mV, ±800mV, ±2V, ±8V, ±20V
Measured potential resolution
0.0008% of potential range, minimum 14nV
Measured potential accuracy
0.2% or 2mV
Impedance Analyser
Frequency range
10µHz to 10kHz
Amplitude
0.15mV to 2.0V, or 0.015% to 100% of current range
DC offset
16bit DC offset subtraction and 2 DC-decoupling filters
Electrometer
Input impedance
>1000Gohm //<10pF
Input bias current
<20pA
Bandwidth
>5MHz
Special functions
Ohmic drop compensation
2V/current range
Peripheral connections
Analog in/out
2/1 (±10V, inputs 18bit resolution, output 16bit resolution, bandwidth 40kHz)
Digital input/output
1 input (0-5V) / 3 output (0-3.3V)
I-out/E-out
Analog monitor for cell current and potential
AC-in
±4V
AC-out
±0.5V sinewave, 10µHz-1MHz with var. att.
Channel-X input, Channel-Y input
±4V: to record impedance from peripheral devices
Environment
Power requirements
100-240V, 50-60Hz, 2000W
Interfacing
USB
Size (w x d x h)
44 x 37 x 8.8cm
PC requirements
Windows 8/10/11, with free USB port
Specifications subject to change, Ivium Technologies ©2026

Model comparison

HE80-1Q
HE80
System performance
Current compliance
0 to +80A
±80A
Maximum applied voltage
+0.5 to +20V
±20V
Potentiostat
Current ranges
+10A and +100A
±10A and ±100A

Need help?

Our experts are ready to answer your questions.

Information request

Postal address

  • Ivium Technologies B.V.
  • De Zaale 11
  • 5612 AJ Eindhoven
  • The Netherlands

Visiting address

  • Ivium Technologies B.V.
  • De Lismortel 31
  • 5612 AR Eindhoven
  • The Netherlands

Contact

  • T +31 40 2390600
  • F
  • E info@ivium.com
©2026 Ivium Technologies B.V.